Quality Product ~ Made in Taiwan
Product Description
- Can measure DC and AC(40-500Hz) magnetic field.
- DC N/S polarity of magnetic field display.
- After the machine components is processed, remain the magnetism to measurement.
- Magnetic application products, the intensity of magnetic field measurement.
- After processing the stress with stainless steel material, remain the measurement of the magnetism
- Magnetic force measurement of the magnetizable material.
- Various steel products brought magnetic measurement naturally.
- For electrical home appliances products, the magnetic field intensity measurement of the motor
- Magnetic field intensity measurement of the permanent magnet.
- Exceed the leakage magnetic field of conducting the magnetite to measure
Application
- DC magnetic field meter measuring static magnetic strength.
- An electromagnetic wave (AC) simply means the wave motion of the electromagnetic field (EMF).
- Measuring extremely low frequency (ELF) of 40 to 500Hz.
- It is capable of measuring the electromagnetic field radiation intensity that is produced by electric transmission equipment, lower lines, microwave ovens, air conditioners, refrigerators, computer monitors, video/audio devices, and so forth.
- The magnetic field unit is Tesla (T), Gauss (G), milli-Gauss (mG), or micro-Tesla (uT).
- 1 T=10,000 G, 1uT=10 mG, 1 G=1,000 mG
Features
- The unit converts: 1mT=10G(Gauss)
- DC N/S polarity of magnetic field display.
- Relative / Peak Hold / Real-time Zero function.
- LCD Back-light: Manually in a back-light to open and close, connect as connecting the power, in a back-light and long and on, it is in a back-light to connect the power and remove to close.
- DATA HOLD.
- MAX/MIN hold.
- GO and NO-GO alarm and alarm value set up the function, Production testing.
- Gauss/Tesla unit selection.
- Manual 200 capacity records.
- Auto 6000 capacity data logging, the interval is set up by PC.
- Maximum and minimum average: record the data downloaded to the computer, automatically out of the calculation of the top 10 maximum / 10 minimum/average.
- Transmission interface: USB interface to send data + PC connection application software, real-time function.
Specifications
- Range: 0~3,000mT(milli Tesla) / 0~30,000G(Gauss), Auto Range
- Resolution: 0.1/1 mG or 0.01/0.1 mT
- AC Frequency response: 40Hz to 500Hz.
- Accuracy at 23 ±3ºC, RH<85%
- DC milli Tesla:
- Range: 300.00mT
- Resolution: 0.01mT
- Accuracy: ±(4% + 10dgt)
- Range: 2,000.0mT
- Resolution: 0.1mT
- Accuracy: ±(4% + 10dgt)
- Range: 2,000.1 ~ 3,000.0mT
- DC Gauss:
- Range: 3,000.0G
- Resolution: 0.1G
- Accuracy: ±(4% + 10dgt)
- Range: 20,000.0G
- Resolution: 1G
- Accuracy: ±(4% + 10dgt)
- Range: 20,001 ~ 30,000G
- AC milli Tesla:
- Range: 150.00mT
- Resolution: 0.01mT
- Accuracy: ±(5% + 20dgt)
- Range: 1,500.0mT
- Resolution: 0.1mT
- Accuracy: ±(5% + 20dgt)
- AC Gauss:
- Range: 1,500.0G
- Resolution: 0.1G
- Accuracy: ±(5% + 20dgt)
- Range: 15,000.0G
- Resolution: 1G
- Accuracy: ±(5% + 20dgt)
- Temperature:
- Range: -20~50℃>/-4~122℉
- Resolution: ±0.1℃/±0.1℉
- Accuracy: ±1.0℃/±1.8℉
- Dimension: 156 x 73 x 35 mm (L x W x H)
- Battery: 6 x AAA batteries
- Operating temperature and humidity: 5℃ ~ 40℃
- External AC 100~240V to DC 9V/0.5A power supply
Standards Certification: