Features
- The liquid level sensor that became easier to use with the output operation switching function and sensitivity switching function.
- Mode setting is adjustable at will with output operation switching function.
- In sensitivity switching function, the transparent, translucent pipe is ø6 up to 13 mm, wall thickness is 1 mm.
- Use clean code (powder without cement).
- Power supply voltage is DC12 to 24V.
Spesification
- UOM : 1.6 x 1.9 x 2.6 CM
- Applicable pipe : Any 6- to 13-mm-diameter pipe with a wall thickness of 1 mm that is made of FEP or any other material as transparent as FEP.
- Sensing object : Liquids in pipes (High-viscosity liquids or liquids with floating materials may not be detected.)
- Light source: GaAs infrared LED with peak wavelength 940 nm
- Indicator: Light indicator GaP (Red LED: Peak wavelength of 700 nm)
- Supply voltage: 12 to 24 VDC ±10%, ripple (p-p): 5% max.
- Current consumption: Average: 30 mA max., Peak: 80 mA max
Control output
- NPN open collector:
- Load power supply voltage: 5 to 24 VDC
- Load current: 100 mA max.
- OFF current: 0.5 mA max.
- 100 mA load current with a residual voltage of 0.8 V max.
- 40 mA load current with a residual voltage of 0.4 V max.
Ambient temperature : 3,000 lx max. with incandescent light or sunlight on the surface of the receiver
Ambient humidity
- Operating: - 10 to +55 ° C
- Storage: - 25 to +65 ° C (with no icing or condensation)
Vibration resistance :Destruction: 10 to 500 Hz, 1.0-mm single amplitude or 150 m/s2 in X, Y, and Z directions 3 times and for 11 min each
Shock resistance: Destruction: 500 m/s2 for 3 times each in X, Y, and Z directions
Degree of protection: IP50
Connection method Pre-wired (Standard length: 1 m)
WeightApprox.: 55 g
Material
Accessories : Support belt (2), slip protection tube (2), Instruction Manual